001
Offsite preparation
Starting with the design of the test, detailed descriptions of techniques and phantoms, calibration and finally the delivery of the equipment.
- Design of E2E Tests in discussion with the Proton Center
- Communication of necessary information to the Proton Center:
- Description of the phantoms for each E2E Test (shape, dimensions, weight, etc.)
- Description of target, dose constraints (if applicable), dose prescription and planning constraints for each E2E test
- Description of the dosimetry techniques used for the E2E tests.
- Delivery of phantoms and equipment Read more
- Calibration and QA of alanine pellets in high-energy proton beam.
- Calibration and QA of EBT3 films in high-energy proton beam.
- Verified calibration of ionization chambers at the NPL.
- Communication of tasks to be performed by the Proton Center and continued support by a MedAustron expert.
- Preparation instruction from a medical physicist and continued support.
002
On-site support
Communication of tasks by MedAustron to be performed by the Proton Center and continued support by a MedAustron expert:
- CT imaging of the phantoms according to local procedures
- Planning of prescription for each selected E2E test according to local procedures
- Phantom immobilization, alignment and positioning according to local procedures
- Delivery of the planned beam delivery sequence according to local procedures
003
Data analysis
Performed by a MedAustron Medical Physicist in cooperation with the NPL:
- Details on the analysis performed can be found in the Audit Report chapter.
004
Audit report
Provision of a final audit report containing:
- Description of the methods of the audit and the measurement equipment used.
- Description of the phantom, target, constraints (if applicable) and dose prescription for each E2E test performed.
- NPL certificate containing:
- Co-60 equivalent absorbed dose to water values for each alanine pellet according to Accredited Radiotherapy Level Alanine Dosimetry service
- Absorbed dose to water values for each alanine pellet corrected for high-energy proton response as measured by MedAustron
- Corrections of the alanine pellets’ measurements provided by NPL for the proton energy dependence.
- Measurement results and comparison with prescribed absorbed dose to water values.
- For each E2E test, results for 22 alanine pellet positions and for one point measurement with Farmer ionization chamber are reported.
- Uncertainty budget of measurement results.